Generic filters

Automatic coagulometer COAG L

Features:

  • Full menu of tests (PT, APTT, TT, fibrinogen, Coagulation Factors, D-dimer, AT-III and inhibitors of blood coagulation).
  • No need for calibration (calibration is performed using a bar code provided with the reagents kit).
  • 4-channel measurement in automatic mode.
  • Special fast mode option (up to 140 tests/hour).
  • System of sampling from open and closed tubes.
  • Data entry through a 17″ color liquid crystal screen.
  • 24/7 live remote access.

Technical parameters:

Throughput

up to 140 tests per hour

Number of measuring channels

4 reading channels

Measurement principle

•        photometry – 405 nm;

•        turbidimetry – 570 nm;

•        nephelometry – 640 nm.

Cuvette incubation positions

60 positions (10 positions in a rack) + 5 STAT positions for urgent analysis

Reagent positions

14 positions + separate positions for solutions

Barcode scanner

built-in; reading information about samples and reagents

Sample and reagent dosage

5-200 μl

  Cuvette number on board

500 pcs

Calibration

automatic (reading information from a barcode scanner) or manual

Quality Control (QC)

special menu for quality control (QC); construction of Levi-Jennings diagrams

Memory

1,000,000 test results and corresponding curves according to identification number (ID), date and time of measurement

Data input

connection to PC and LIS

Size

85 х 75 х 51 cm

Weight

80 kg



Main advantages:

  • Full menu of tests (PT, APTT, TT, fibrinogen, blood coagulation factors, D-dimer, AT-III and inhibitors of blood coagulation).
  • No need for calibration (calibration is performed using a bar code provided in the kit with reagents).
  • 4-channel measurement in automatic mode.
  • Special fast mode option (up to 140 PT tests/hour).
  • System of sampling from open and closed tubes.
  • Data entry through a 17″ color liquid crystal screen.
  • Real-time remote access 24/7.

 

Technical parameters:

Throughput up to 140 PT tests per hour
Number of measuring channels 4 reading channels
 

Measurement principle

•        photometry – 405 nm;

•        turbidimetry – 570 nm;

•        nephelometry – 640 nm.

Cuvette incubation positions 60 positions (10 positions in a rack) + 5 STAT positions for urgent analysis
Reagent positions 14 positions + separate positions for solutions
Barcode scanner built-in; reading information about samples and reagents
Sample and reagent dosage 5-200 μl
  Cuvette number on board 500 pcs
Calibration automatic (reading information using a barcode scanner) or manual
Quality Control (QC) special menu for quality control (QC); construction of Levi-Jennings charts
Memory 1,000,000 test results and corresponding curves according to identification number (ID), date and time of measurement
Data input connection to PC and LIS
Size 85 х 75 х 51 cm
Weight 80 kg

 